Skip to main navigation menu Skip to main content Skip to site footer

Characterization and Commissioning of a Triple-GEM Detector

Abstract

We present in detail the characterization and commissioning of a triple Gas Electron Multiplier (triple-GEM)
detector. The response to various sources of radiation was studied, and its efficiency, gain, energy resolution,
and temporal resolution were calculated. We found an energy resolution of 19.5%, a temporal resolution of
40 ns., and a maximum gain of 5×106. We conclude that the detector can be confidently used in further
studies of medical dosimetry, since it is working in the proportional mode.

Keywords

Commissioning, Detector characterization, GEM detector, Particle detectors, Radioactivity, X-rays

PDF (Español)

Author Biography

Héctor Fabio Castro Serrato

https://scholar.google.com/citations?user=wC2dY4sAAAAJ&hl=es


References

  • L. Guirl, S. Kane, J. May, J. Miyamoto, and I. Shipsey, “An aging study of triple gems in Ar-CO2,” Nuclear Instruments and Meth- ods in Physics Research Section A: Accelera- tors, Spectrometers, Detectors and Associated Equipment, vol. 478, no. 1-2, pp. 263–266, 2002. DOI: https://doi.org/10.1016/S0168-9002(01)01768-5
  • B. Ketzer, M. Altunbas, K. Dehmelt, J. Ehlers, J. Friedrich, B. Grube, S. Kappler, I. Konorov, S. Paul, A. Placci, et al., Triple gem tracking detectors for compass, IEEE Transactions on Nuclear Science 49 (5) (2002) 2403–2410. DOI: https://doi.org/10.1109/TNS.2002.803891
  • S. Lami, G. Latino, E. Oliveri, L. Ropelewski, N. Turini, T. Collaboration, et al., A triple- gem telescope for the totem experiment, Nu- clear Physics B-Proceedings Supplements 172 (2007) 231–233. DOI: https://doi.org/10.1016/j.nuclphysbps.2007.08.055
  • M. Poli Lener, L.-G. group, Triple-gem detec- tors for the innermost region of the lhcb muon apparatus, in: AIP Conference Proceedings, Vol. 794, American Institute of Physics, 2005, pp. 311–314. DOI: https://doi.org/10.1063/1.2125678
  • D. Abbaneo, M. Abbrescia, M. Abi Akl, C. Armaingaud, P. Aspell, Y. Assran, S. Bally, Y. Ban, P. Barria, L. Benussi, et al., Status of the triple-gem project for the upgrade of the cms muon system, Journal of Instrumentation 8 (12) (2013) C12031. DOI: https://doi.org/10.1088/1748-0221/8/12/C12031
  • G. D. D. CERN, https://gdd.web.cern.ch/, Gas Detectors Development Group at CERN.
  • T. Behnke, R. Diener, C. Rosemann, L. Steder, A novel self-supporting gem-based amplifica- tion structure for a time projection chamber at the ilc, Journal of Instrumentation 8 (12) (2013) P12009. DOI: https://doi.org/10.1088/1748-0221/8/12/P12009
  • F. Simon, J. Kelsey, M. Kohl, R. Ma- jka, M. Plesko, D. Underwood, T. Sakuma, N. Smirnov, H. Spinka, B. Surrow, Triple gem detectors for the forward tracker in star, in: 2007 IEEE Nuclear Science Symposium Con- ference Record, Vol. 1, IEEE, 2007, pp. 234– 238. DOI: https://doi.org/10.1109/NSSMIC.2007.4436321
  • B. Azmoun, N. Smirnov, S. Stoll, C. Woody, Test of a gem detector in the phenix experiment at rhic, in: IEEE Symposium Conference Record Nuclear Science 2004., Vol. 1, IEEE, 2004, pp. 480–484.
  • A. Velásquez and H. Castro, “Characterization and calibration of a Triple-GEM detector for medical dosimetry,” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 1000, p. 165241, 2021. DOI: https://doi.org/10.1016/j.nima.2021.165241
  • D. N. Poenaru, W. Greiner, Experimental techniques in nuclear physics, Walter de Gruyter, 2011.
  • W. Leo, Techniques for nuclear and particle physics experiments: a how-to approach. Springer Science & Business Media, 2012.
  • D. Coy, “Construcción y caracterización de un detector de partículas tipo resistive plate cham- ber,” Master’s thesis, Universidad Nacional de Colombia, 2014.
  • E. Rozo, “Desarrollo de un detector de rayos X y electrones para uso en radioterapia,” Master’s thesis, Universidad Nacional de Colombia, 2010.
  • F. Murtas, “Development of a gaseous detector based on gas electron multiplier (GEM) tech- nology,” 2002, national Laboratory of Frascati, Istituto Nazionale di Fisica Nucleare.
  • F. Sauli, “GEM: A new concept for electron amplification in gas detectors,” Nuclear Instru- ments and Methods in Physics Research Sec- tion A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 386, no. 2-3, pp. 531–534, 1997. DOI: https://doi.org/10.1016/S0168-9002(96)01172-2
  • F. Sauli, “The gas electron multiplier (GEM): Operating principles and applications,” Nu- clear Instruments and Methods in Physics Re- search Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 805, pp. 2–24, 2016. DOI: https://doi.org/10.1016/j.nima.2015.07.060
  • J. Wang, Upgrade plans and ageing studies for the cms muon system in preparation of hl-lhc, arXiv preprint arXiv:1809.01760 (2018).
  • J. Useche, C. Avila, Estimation of cosmic- muon flux attenuation by monserrate hill in bogotá, arXiv preprint arXiv:1810.04712 (2018).
  • B. Dorney, “Gas detector physics: Principles and fundamentals,” May 2018.
  • L. Cruz, “Caracterización y prueba de la cámara de ionización Triple-GEM a partir de muones resultantes de rayos cósmicos,” Master’s thesis, Universidad de los Andes, 2014.
  • R. N. Patra, R. N. Singaraju, S. Biswas, Z. Ahammed, T. K. Nayak, and Y. P. Viyogi,
  • “Measurement of basic characteristics and gain uniformity of a triple GEM detector,” Nuclear
  • Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detec- tors and Associated Equipment, vol. 862, pp. 25–30, 2017. DOI: https://doi.org/10.1016/j.nima.2017.05.011
  • M. Zecchin, “Characterization of a Triple- GEM detector prototype for the CMS muon spectrometer upgrade with gem detectors,” Tech. Rep., 2014.
  • S.Han,Y.Kim,J.M.Han,Y.S.Cho,Y.J.Kim, K. Jeong, H. Rang, D. H. Kim, H. Cho, S. Kang et al., “Performance of a double GEM and a triple GEM,” Journal of the Korean Physical Society, vol. 42, no. 5, pp. 606–611, 2003.

Downloads